Call us on 01707 385 226

Medin CNO®

medinCNO® – Non-invasive respiratory support for preterm and newborn babies

  • nCPAP with nHFV
  • Synchronized NIPPV
  • Apnea monitoring
  • MediTRIG: Integrated trigger without additional sensor
  • Leakage compensation

Comprehensive non-invasive respiratory support

medinCNO is a versatile nCPAP device which is used to provide respiratory support to premature infants and newborns. In combination with Medijet®, the compact device is a stand-alone system and meets all requirements for non-invasive nCPAP therapy.

With five modes, medinCNO offers comprehensive non-invasive support in neonatology and is the only nCPAP device with nHFV.

Features

  • nCPAP mode

When using nCPAP, the patient must be able to breathe spontaneously, since no breaths are administered. The continuous positive pressure ensures stabilization of the airways and lungs.

nCPAP is the most commonly used mode of non-invasive respiratory support in premature infants and newborns and is the basis of some further variants, such as NIPPV, SNIPPV and nHFV.

  • ApneaCPAP mode

Apnea is detected by MediTRIG. As a result, breaths are automatically triggered to stimulate breathing. Frequency, pressure level and breath duration can be individually adjusted.

  • NIPPV

In addition to the PEEP, breaths with fixed settings (pressure level, duration, frequency) are administered. These breaths have a higher pressure level and are not emitted in a synchronized manner.

  • SNIPPV

Mode which is synchronized with the patient’s spontaneous breathing with the aid of MediTRIG. With his/her inspiration, the patient can trigger an assisted breath which supports inspiration with flow and pressure.

  • nHFO – nasal High Frequency Ventilation

nHFV combines the benefits of high-frequency ventilation and non-invasive CPAP support. The patient breathes spontaneously with CPAP support. A special valve shifts the air flow in all phases of spontaneous breathing under high-frequency oscillations (300 to 1200/minute).

  • MediTRIG

The MediTRIG respiratory trigger was specially developed by medin for use in premature infants and newborns. Through the combination of the pressure measurement in the Medijet® generator and the signal processing in the medin nCPAP devices, MediTRIG works without additional sensors.

MediTRIG detects central as well as obstructive apnea.

Important note – Some features may be available as options, may have different specifications than described on this website, or may not be available. Contact your medin representative for details.

 

Please call our Customer Care team to discuss your requirements on 01707 333164 or

Download the Medin product catalogue M0137 01 Medin Product Catalogue Nov 2017